FreeCalypso > hg > freecalypso-docs
comparison CMU200-maintenance-notes @ 83:a87d9ee278fb
CMU200-maintenance-notes: new article
| author | Mychaela Falconia <falcon@freecalypso.org> |
|---|---|
| date | Thu, 13 Jan 2022 08:18:03 +0000 |
| parents | |
| children | d2fef140ed53 |
comparison
equal
deleted
inserted
replaced
| 82:e279a924f7a3 | 83:a87d9ee278fb |
|---|---|
| 1 Rohde & Schwarz CMU200 instrument is an absolutely essential piece of test | |
| 2 equipment for anyone in the business (or hobby) of designing and building his | |
| 3 or her own personal cellphones of 2G and/or 3G variety. I (Mother Mychaela) | |
| 4 currently only work with GSM, but depending on installed hw and sw options, | |
| 5 CMU200 instruments also support AMPS, IS-136, IS-95 (CDMA 2G) and both WCDMA | |
| 6 and CDMA2000 varieties of 3G. | |
| 7 | |
| 8 Over the course of owning and maintaining a CMU200 instrument since 2017 and | |
| 9 having had to repair it twice now (as of 2022-01), and having conversed with | |
| 10 another CMU200 owner who had to repair his instrument in the same way, I | |
| 11 started observing a pattern in that many of these instruments are now failing | |
| 12 in the field in exactly the same ways. All of these failures happen in the | |
| 13 RXTX board, and the purpose of this article is to educate other CMU instrument | |
| 14 owners about these failures and most importantly, how to repair them. | |
| 15 | |
| 16 Credit attribution | |
| 17 ================== | |
| 18 | |
| 19 I sincerely thank Michael Katzmann, NV3Z / VK2BEA / G4NYV, for his invaluable | |
| 20 help in reverse-engineering the insides of the culprit RXTX board, identifying | |
| 21 various critical components on that board, including the ones that habitally | |
| 22 fail, and identifying Eccosorb-caused galvanic corrosion as the root cause of | |
| 23 these failures. Without his help, I would not have made it this far! | |
| 24 | |
| 25 What is this RXTX board | |
| 26 ======================= | |
| 27 | |
| 28 This board is common among CMU200, CMU300 and CRTU-RU instruments from R&S - or | |
| 29 at least these are the ones I know - maybe there are others I don't know about. | |
| 30 This board encapsulates the instrument's main RF Rx and Tx chains: on the Rx | |
| 31 side it takes RF input from the front end and performs triple IF downconversion | |
| 32 to 10.7 MHz IF3, and on the Tx side it takes 13.85 MHz IF3 input and upconverts | |
| 33 it to RF output, going through IF2 and IF1 in the process - triple IF in both | |
| 34 directions. | |
| 35 | |
| 36 Every CMU200 instrument always has one RXTX board - it is an absolutely required | |
| 37 component irrespective of option configurations. The hardware architecture of | |
| 38 this instrument also has a place for an optional second RXTX board, providing a | |
| 39 complete second Rx and Tx channel - however, as far as I can tell, CMU200 | |
| 40 software won't do anything with it, i.e., there are no test modes or | |
| 41 applications in CMU200 software repertoire that can make use of a second RXTX | |
| 42 board. Instead it seems that configurations with two RXTX boards are better | |
| 43 supported on the CRTU-RU platform - but I know next to nothing about that one. | |
| 44 | |
| 45 Also note: if your CMU200 is equipped with Aux Tx model B96 (as opposed to B95), | |
| 46 there is an output from that B96 add-on that goes to the front end input that | |
| 47 was originally meant for second RXTX. | |
| 48 | |
| 49 RXTX board failures | |
| 50 =================== | |
| 51 | |
| 52 In terms of externally visible symptoms, almost all CMU200 units are now failing | |
| 53 in the same ways: | |
| 54 | |
| 55 1) If Tx side fails, the visible symptom is completely absent or extremely weak | |
| 56 output, and the internal loopback test fails with no signal detected at any of | |
| 57 the frequencies in the test sequence. A key point is that this failure mode is | |
| 58 independent of the selected output frequency. | |
| 59 | |
| 60 2) If Rx side fails, different frequency ranges are affected differently. As I | |
| 61 shall explain momentarily, there are two different IF1 Rx paths inside the RXTX | |
| 62 board: one handles the frequency range from > 1200 to <= 2200 MHz, and the other | |
| 63 handles lower (<= 1200 MHz) and higher (> 2200 MHz) input frequencies. When a | |
| 64 given RXTX board develops Rx path failure, this failure happens separately in | |
| 65 each of these two IF1 Rx paths. The resulting symptoms vary: if only one of | |
| 66 the two IF1 Rx paths fails, then only that frequency range will be affected, | |
| 67 or if both fail, the observed loss will typically be different between the two | |
| 68 frequency ranges. The failure symptom is unexpected large attenuation: | |
| 69 sometimes around 5 to 6 dB of loss, othertimes as much as 25 dB of loss. | |
| 70 | |
| 71 RXTX board architecture explained | |
| 72 ================================= | |
| 73 | |
| 74 Unfortunately R&S' official service manual for CMU200 instruments is only a part | |
| 75 swapper guide: it tells you which boards do what in general terms and tells you | |
| 76 how to remove and replace each part, but no schematics, and no detailed | |
| 77 explanation of what happens inside each board. However, I draw the reader to | |
| 78 the block diagram on page 3.2 of this manual - this block diagram does provide | |
| 79 an important starting point for understanding what happens inside the RXTX | |
| 80 board. | |
| 81 | |
| 82 In the Tx direction, 13.85 MHz IF3 comes in from the digital board - or from | |
| 83 B68 board in WCDMA test modes. This Tx IF3 is mixed with Tx LO3 to produce | |
| 84 Tx IF2. This Tx IF2 is fixed at 487.52 MHz, thus one would think that Tx LO3 | |
| 85 frequency ought to be fixed as well - but it seems to be a synthesized variable | |
| 86 frequency. (Remember, all of this understanding is from reverse engineering, | |
| 87 hence we can only figure out so much.) | |
| 88 | |
| 89 Tx IF2 of 487.52 MHz is then passed through a pair of identical SAW filters, | |
| 90 Sawtek 855272 - two cascaded identical filters, with an amplifier in between. | |
| 91 This SAW filter has a center frequency of 479.75 MHz with 20 MHz bandwidth, | |
| 92 thus the passband spans from 469.75 to 489.75 MHz. Notice how Tx IF2 of | |
| 93 487.52 MHz stands just 2.23 MHz away from the edge of the passband - is it | |
| 94 intentional? What are they filtering? Without original design notes, we can | |
| 95 only guess. As I shall explain later in this article, one of these two Tx IF2 | |
| 96 SAW filters is a component prone to failure. | |
| 97 | |
| 98 After these cascaded SAW filters, Tx IF2 is mixed with LO2. Unlike LO1 and LO3, | |
| 99 there is only one LO2 for both Rx and Tx, and it is fixed at 1329.6 MHz. When | |
| 100 Tx IF2 at fixed 487.52 MHz is mixed with LO2 at fixed 1329.6 MHz, the output of | |
| 101 this mixer will always contain two frequencies: 842.08 MHz and 1817.12 MHz. | |
| 102 These are the two possible Tx IF1 frequencies, and there is a frequency- | |
| 103 selective filter for each of these two Tx IF1 modes. Based on the final output | |
| 104 frequency to be generated, instrument control software selects either low or | |
| 105 high Tx IF1, controlling switches before and/or after the filters. I have not | |
| 106 investigated to see if the frequency ranges for high vs. low Tx IF1 are the same | |
| 107 as on the Rx side or not - maybe they are the same, maybe they are different. | |
| 108 | |
| 109 After Tx IF1 output is combined or switched from the two filters, it is mixed | |
| 110 with Tx LO1 to produce the final RF output. The mixer that does this job is | |
| 111 MACOM SM4T, which is one of the larger, prominently visible components on the | |
| 112 board. There also seems to be a fourth mixer and LO stage that kicks in only | |
| 113 for frequencies above 2200 MHz, but I haven't really studied that one as my main | |
| 114 interest is in the classic cellular frequency bands, 1900 MHz and below. | |
| 115 | |
| 116 On the Rx side the same process happens in reverse, but the specific frequencies | |
| 117 used for IF1, IF2 and IF3 are slightly different. At first there is a stage | |
| 118 that only kicks in for frequencies above 2200 MHz (bypassed otherwise), and | |
| 119 then there is an SM4T mixer (identical to the one on Tx side) that takes in RF | |
| 120 and Rx LO1 to produce Rx IF1. High-side injection is used, i.e., Rx LO1 is | |
| 121 programmed to generate frequency equal to the external RF of interest PLUS the | |
| 122 desired Rx IF1 output. | |
| 123 | |
| 124 Rx LO1 is programmed as follows by the instrument control software: | |
| 125 | |
| 126 * Rx IF1 will be at 1816.115 MHz (call it high) if the listening frequency is | |
| 127 <= 1200 MHz or > 2200 MHz; | |
| 128 | |
| 129 * Rx IF1 will be at 843.085 MHz (call it low) if the listening frequency is in | |
| 130 the intermediate range, i.e., 1200 MHz < RF <= 2200 MHz. | |
| 131 | |
| 132 In addition to programming Rx LO1 to produce the desired IF1 per the logic | |
| 133 above, the software also controls switches that select one or the other IF1 | |
| 134 filter: either the filter that passes low IF1 or the one that passes high IF1. | |
| 135 | |
| 136 The filters used for low and high IF1 modes are the same on both Rx and Tx | |
| 137 sides. (The actual frequencies are slightly different, but in each case they | |
| 138 fit within the passband of the common filter parts.) The filter for low IF1 is | |
| 139 Murata DFC3R836P025HHD, package marking 836 CD, and the one for high IF1 is | |
| 140 DFC31R84P075HHA, package marking CR. The two filter packages are NOT the same | |
| 141 mechanically: the low IF1 filter is physically larger. Both parts are ceramic | |
| 142 monoblock filters from the same family, and it seems that these filter parts | |
| 143 were originally made for mobile phones, not for RF metrology instruments: the | |
| 144 "836 CD" filter is for AMPS uplink band, and the "CR" filter is for DCS downlink | |
| 145 band. | |
| 146 | |
| 147 On the Tx side of the board there are only two IF1 filters: one for low Tx IF1 | |
| 148 and one for high Tx IF1. However, on the Rx side there are 3 of these ceramic | |
| 149 filters in total: two for high IF1 (two cascaded identical filters with an | |
| 150 amplifier in between) and just one for low IF1. Why am I covering these filters | |
| 151 in so much detail? You probably guessed it: they are components that fail, as | |
| 152 will be covered shortly. | |
| 153 | |
| 154 After the selection of either low or high IF1 filter, Rx IF1 coming out of the | |
| 155 selected filter (either 843.085 MHz or 1816.115 MHz) is mixed with LO2, which is | |
| 156 shared between Rx and Tx sides and fixed at 1329.6 MHz. The output of this | |
| 157 mixer is Rx IF2 at 486.515 MHz. This Rx IF2 then passes through a pair of | |
| 158 cascaded Sawtek 855272 filters, two identical filters with an amplifier in | |
| 159 between, exactly the same as on the Tx side. Then there is Rx LO3 and the final | |
| 160 mixer, producing Rx IF3 at 10.7 MHz that goes to the digital board, to the rear | |
| 161 panel BNC output and to the WCDMA board (B68) if the latter is present. | |
| 162 | |
| 163 How these RXTX boards fail | |
| 164 ========================== | |
| 165 | |
| 166 There are 3 specific components on this RXTX board that have been seen to fail | |
| 167 over and over in the field: | |
| 168 | |
| 169 * The second of the two cascaded IF2 SAW filters (Sawtek 855272) on the Tx side | |
| 170 often fails, breaking the Tx chain (output totally gone or extremely weak) | |
| 171 for all frequencies. Note that there are a total of 4 identical Sawtek 855272 | |
| 172 filters on this board (2 on Rx side, 2 on Tx side), and only one of the four | |
| 173 fails: Tx side, second filter in the cascade. | |
| 174 | |
| 175 * The "836 CD" filter on the Rx side is prone to failure. When it fails, the | |
| 176 visible symptom is severe attenuation in measured Rx signal levels for input | |
| 177 frequencies in the 1200 MHz < RF <= 2200 MHz range. Only the Rx side filter | |
| 178 fails, not the identical one on the Tx side! | |
| 179 | |
| 180 * One of the two cascaded "CR" filters on the Rx side likewise fails - this time | |
| 181 it is the first one in the cascade. The other two identical "CR" filters on | |
| 182 the same board (the second in cascade for Rx and the one for Tx) are likewise | |
| 183 NOT seen to fail. | |
| 184 | |
| 185 The root cause of all 3 component failures has been traced to galvanic corrosion | |
| 186 caused by direct contact between these components and Eccosorb RF absorber foam. | |
| 187 The complete RXTX board assembly consists of the traditional PCBA plus heavy | |
| 188 metal shields on both sides; the front and back metal shield pieces are custom- | |
| 189 made for this board, with individually shielded cavities matching different | |
| 190 sections of the board. Some (not all) of these cavities are filled with a | |
| 191 special black foam called Eccosorb - it is an RF absorber, presumably added to | |
| 192 prevent these cavities from acting as parasitic oscillators. Trouble occurs | |
| 193 when this Eccosorb foam comes into direct contact with metal surfaces of | |
| 194 components on the board: the result is galvanic corrosion, a process that takes | |
| 195 many years before it results in component failure. The reason why only 3 | |
| 196 particular filter components fail is because they got the bad luck of residing | |
| 197 in cavities with Eccosorb - the other identical components that don't fail | |
| 198 reside in cavities without Eccosorb. | |
| 199 | |
| 200 We don't know how R&S allowed this design flaw to escape and remain in their | |
| 201 sold and field-deployed products: there is the "innocent" explanation that they | |
| 202 simply didn't notice, and there is the conspiratorial view that this slow | |
| 203 failure mechanism is intentional as in planned obsolescense - pick your choice | |
| 204 of hypothesis. | |
| 205 | |
| 206 How to repair failed boards | |
| 207 =========================== | |
| 208 | |
| 209 All 3 of the failing filter components (one SAW filter part and two ceramic | |
| 210 monoblock filter parts) are now unobtainium. However, because so many of these | |
| 211 RXTX boards fail in exactly the same ways, our community at large is now | |
| 212 accumulating a very substantial "graveyard" of failed boards, and here is the | |
| 213 good news: we can make one good board out of every two failed ones. Suppose | |
| 214 that every RXTX board in our community's collective inventory has fully failed, | |
| 215 leaving no failure-free boards - what now? Here is the recipe for making one | |
| 216 good RXTX board out of two fully failed ones: | |
| 217 | |
| 218 1) Out of the two failed boards, choose one to be the part donor and the other | |
| 219 to be the part recipient. | |
| 220 | |
| 221 2) Take the part donor board and harvest 3 parts from it: one of the 3 Sawtek | |
| 222 855272 filters that aren't subject to corrosion, and the two IF1 filters | |
| 223 (one 836 CD and one CR) from the Tx side. Tx side IF1 filters aren't in | |
| 224 contact with Eccosorb and thus don't corrode, and 3 out of the 4 SAW filters | |
| 225 are likewise safe - hence we expect that every "dead" RXTX board can still | |
| 226 serve as a donor of good parts in this manner. | |
| 227 | |
| 228 3) Take the part recipient board and transplant the donor parts onto it, | |
| 229 replacing all 3 corroded filters. | |
| 230 | |
| 231 4) Before putting the repaired board back into its metal casing, cover all | |
| 232 corrosion-prone components with Kapton tape, preventing direct galvanic | |
| 233 contact with Eccosorb - this way the newly transplanted uncorroded components | |
| 234 won't suffer the same fate. | |
| 235 | |
| 236 RXTX disassembly instructions | |
| 237 ============================= | |
| 238 | |
| 239 Before you can start working on an individual RXTX board, you first need to pull | |
| 240 it out of your CMU. Disassembly instructions are provided in the official part | |
| 241 swapper guide from R&S (which they call "service manual"), but here is the gist: | |
| 242 | |
| 243 * Using a Torx T20 screwdriver, remove the 4 rear feet and lift the sleeve part | |
| 244 of the instrument case. | |
| 245 | |
| 246 * Remove two small Phillips screws that secure the cover over the main board | |
| 247 cage, and lift that cover off. | |
| 248 | |
| 249 * Unhook all MMCX little coax connections from the RXTX board: 3 on the top side | |
| 250 (IF3 interface) and one on the bottom (netclock input). | |
| 251 | |
| 252 * Loosen and remove the two semi-rigid coax pieces that connect RF between the | |
| 253 RXTX board and the front end. In this Mother's opinion, this step is the | |
| 254 least pleasant of all, but it is unavoidable. | |
| 255 | |
| 256 * After ensuring that nothing remains connected to the RXTX board on the bottom | |
| 257 side, pull the board out from the top. | |
| 258 | |
| 259 Once you got the complete RXTX board assembly out, how do you extract the actual | |
| 260 board out of the metal casing? The not-immediately-obvious answer is that you | |
| 261 don't need to remove all of the screws, instead there are shortcuts that will | |
| 262 save you a lot of pain: | |
| 263 | |
| 264 * There are two smooth thin metal plates, one on the front side of the board | |
| 265 (facing toward the front of the CMU when installed) and one on the back side. | |
| 266 Each is secured with a small Phillips screw. You only need to remove the one | |
| 267 on the front side. You don't need to remove the thin metal plate from the | |
| 268 back side of RXTX assembly - doing so will only add more clutter and loose | |
| 269 parts to your lab bench while the board is disassembled. | |
| 270 | |
| 271 * Once you remove the thin metal plate from the *front* side of your RXTX | |
| 272 assembly, you will see all of the many screws that hold together the sandwich | |
| 273 of two heavy metal pieces with the board in the middle. These screws are | |
| 274 Torx T8. | |
| 275 | |
| 276 * Put the board down on your bench so that the side that faces the front of the | |
| 277 CMU when installed (the side with the T8 screw heads) will become the top, | |
| 278 with the rear side becoming bottom. | |
| 279 | |
| 280 * Each of the T8 screws passes through thread in the top metal piece, a hole in | |
| 281 the PCB, and then thread in the bottom metal piece. As you loosen these | |
| 282 screws, you don't need to remove them all the way - instead loosen each screw | |
| 283 so that its far end comes out of the thread in the bottom metal piece, but | |
| 284 let it remain captive in the top metal piece. Letting the screws remain | |
| 285 captive in the top metal piece will reduce bench clutter while the board is | |
| 286 disassembled, and there is a lot less screwing and unscrewing work to be done, | |
| 287 as there is no need to work through the thread in the top metal piece. | |
| 288 | |
| 289 Once you loosen all of the T8 screws, the top metal piece should lift off, | |
| 290 leaving just the bottom metal piece and the PCBA. The bottom metal piece has | |
| 291 two thin metal pins sticking out of it; both the PCBA and the top metal piece | |
| 292 align on these two pins. | |
| 293 | |
| 294 When you lift the top metal piece (the one with the screws), the side of the | |
| 295 board that will be immediately exposed to you is the side that faces the front | |
| 296 of the CMU when the board is installed. It is the Rx side, and you can confirm | |
| 297 that you are looking at the Rx side by noting that there are two "CR" filters | |
| 298 for high IF1, as opposed to just one. And chances are, right here at this step | |
| 299 in the disassembly process you will see the galvanic corrosion or the lead-up | |
| 300 to it. | |
| 301 | |
| 302 As you lift the top metal piece from the board, look at its inside and note the | |
| 303 many individual cavities. Also note how some of these cavities are filled with | |
| 304 some black foam - that's the Eccosorb. And note how only some of the cavities | |
| 305 have Eccosorb in them, not all. | |
| 306 | |
| 307 Now look at the ceramic IF1 filters on the Rx side of the board. The one "CR" | |
| 308 filter that is NOT in contact with Eccosorb will be bright copper-colored (it | |
| 309 actually is copper), whereas the two filters that are in contact with Eccosorb | |
| 310 (one 836 CD, one CR) will often be green instead of copper-colored on their top | |
| 311 surface - that's patinated copper! Furthermore, there will typically be some | |
| 312 black Eccosorb material directly adhered to the corroding top surfaces of those | |
| 313 two unlucky filters. | |
| 314 | |
| 315 Now lift the PCBA off the two metal pins, separating it from the bottom metal | |
| 316 piece. Like you did with the top metal piece, observe the inside of the bottom | |
| 317 metal piece: note which cavities have Eccosorb in them and which don't. Then | |
| 318 flip the board over and look at its Tx side. You will see that there are only | |
| 319 two ceramic IF1 filters on this side (one 836 CD and one CR), and both should | |
| 320 be in pristine shape, bright copper-colored, no corrosion - these two are not | |
| 321 in contact with Eccosorb! | |
| 322 | |
| 323 Now look at the two Sawtek 855272 filters on the Tx side. The one closer to | |
| 324 the middle of the board will often appear in worse physical condition that the | |
| 325 other 3 - and the culprit is once again in contact with Eccosorb. | |
| 326 | |
| 327 MACOM SM4T mixer corrosion | |
| 328 ========================== | |
| 329 | |
| 330 Neither I nor my collaborator on this project have seen an RXTX board on which | |
| 331 either the Rx SM4T mixer or the Tx one went bad - i.e., we haven't seen a | |
| 332 failure in this part *yet*. However, this mixer *is* in contact with Eccosorb, | |
| 333 and looking visually at the collection of RXTX boards in my possession, I | |
| 334 (Mother Mychaela) see definite signs of corrosion - the metal surface of this | |
| 335 SM4T mixer component is beginning to corrode. Therefore, as a preventative | |
| 336 measure, I recommend cleaning off any Eccosorb that is adhered to this component | |
| 337 and then covering the component with Kapton tape before putting the board back | |
| 338 into its metal casing. | |
| 339 | |
| 340 Unlike the failing filters, this MACOM SM4T mixer is still available new - but | |
| 341 it's an expensive component, so let's protect these mixers from corrosion. |