Rohde & Schwarz CMU200 instrument is an absolutely essential piece of test equipment for anyone in the business (or hobby) of designing and building his or her own personal cellphones of 2G and/or 3G variety. I (Mother Mychaela) currently only work with GSM, but depending on installed hw and sw options, CMU200 instruments also support AMPS, IS-136, IS-95 (CDMA 2G) and both WCDMA and CDMA2000 varieties of 3G. Over the course of owning and maintaining a CMU200 instrument since 2017 and having had to repair it twice now (as of 2022-01), and having conversed with another CMU200 owner who had to repair his instrument in the same way, I started observing a pattern in that many of these instruments are now failing in the field in exactly the same ways. All of these failures happen in the RXTX board, and the purpose of this article is to educate other CMU instrument owners about these failures and most importantly, how to repair them. Credit attribution ================== I sincerely thank Michael Katzmann, NV3Z / VK2BEA / G4NYV, for his invaluable help in reverse-engineering the insides of the culprit RXTX board, identifying various critical components on that board, including the ones that habitually fail, and identifying Eccosorb-caused galvanic corrosion as the root cause of these failures. Without his help, I would not have made it this far! What is this RXTX board ======================= This board is common among CMU200, CMU300 and CRTU-RU instruments from R&S - or at least these are the ones I know - maybe there are others I don't know about. This board encapsulates the instrument's main RF Rx and Tx chains: on the Rx side it takes RF input from the front end and performs triple IF downconversion to 10.7 MHz IF3, and on the Tx side it takes 13.85 MHz IF3 input and upconverts it to RF output, going through IF2 and IF1 in the process - triple IF in both directions. Every CMU200 instrument always has one RXTX board - it is an absolutely required component irrespective of option configurations. The hardware architecture of this instrument also has a place for an optional second RXTX board, providing a complete second Rx and Tx channel - however, as far as I can tell, CMU200 software won't do anything with it, i.e., there are no test modes or applications in CMU200 software repertoire that can make use of a second RXTX board. Instead it seems that configurations with two RXTX boards are better supported on the CRTU-RU platform - but I know next to nothing about that one. Also note: if your CMU200 is equipped with Aux Tx model B96 (as opposed to B95), there is an output from that B96 add-on that goes to the front end input that was originally meant for second RXTX. RXTX board failures =================== In terms of externally visible symptoms, almost all CMU200 units are now failing in the same ways: 1) If Tx side fails, the visible symptom is completely absent or extremely weak output, and the internal loopback test fails with no signal detected at any of the frequencies in the test sequence. A key point is that this failure mode is independent of the selected output frequency. 2) If Rx side fails, different frequency ranges are affected differently. As I shall explain momentarily, there are two different IF1 Rx paths inside the RXTX board: one handles the frequency range from > 1200 to <= 2200 MHz, and the other handles lower (<= 1200 MHz) and higher (> 2200 MHz) input frequencies. When a given RXTX board develops Rx path failure, this failure happens separately in each of these two IF1 Rx paths. The resulting symptoms vary: if only one of the two IF1 Rx paths fails, then only that frequency range will be affected, or if both fail, the observed loss will typically be different between the two frequency ranges. The failure symptom is unexpected large attenuation: sometimes around 5 to 6 dB of loss, othertimes as much as 25 dB of loss. RXTX board architecture explained ================================= Unfortunately R&S' official service manual for CMU200 instruments is only a part swapper guide: it tells you which boards do what in general terms and tells you how to remove and replace each part, but no schematics, and no detailed explanation of what happens inside each board. However, I draw the reader to the block diagram on page 3.2 of this manual - this block diagram does provide an important starting point for understanding what happens inside the RXTX board. In the Tx direction, 13.85 MHz IF3 comes in from the digital board - or from B68 board in WCDMA test modes. This Tx IF3 is mixed with Tx LO3 to produce Tx IF2. This Tx IF2 is fixed at 487.52 MHz, thus one would think that Tx LO3 frequency ought to be fixed as well - but it seems to be a synthesized variable frequency. (Remember, all of this understanding is from reverse engineering, hence we can only figure out so much.) Tx IF2 of 487.52 MHz is then passed through a pair of identical SAW filters, Sawtek 855272 - two cascaded identical filters, with an amplifier in between. This SAW filter has a center frequency of 479.75 MHz with 20 MHz bandwidth, thus the passband spans from 469.75 to 489.75 MHz. Notice how Tx IF2 of 487.52 MHz stands just 2.23 MHz away from the edge of the passband - is it intentional? What are they filtering? Without original design notes, we can only guess. As I shall explain later in this article, one of these two Tx IF2 SAW filters is a component prone to failure. After these cascaded SAW filters, Tx IF2 is mixed with LO2. Unlike LO1 and LO3, there is only one LO2 for both Rx and Tx, and it is fixed at 1329.6 MHz. When Tx IF2 at fixed 487.52 MHz is mixed with LO2 at fixed 1329.6 MHz, the output of this mixer will always contain two frequencies: 842.08 MHz and 1817.12 MHz. These are the two possible Tx IF1 frequencies, and there is a frequency- selective filter for each of these two Tx IF1 modes. Based on the final output frequency to be generated, instrument control software selects either low or high Tx IF1, controlling switches before and/or after the filters. I have not investigated to see if the frequency ranges for high vs. low Tx IF1 are the same as on the Rx side or not - maybe they are the same, maybe they are different. After Tx IF1 output is combined or switched from the two filters, it is mixed with Tx LO1 to produce the final RF output. The mixer that does this job is MACOM SM4T, which is one of the larger, prominently visible components on the board. There also seems to be a fourth mixer and LO stage that kicks in only for frequencies above 2200 MHz, but I haven't really studied that one as my main interest is in the classic cellular frequency bands, 1900 MHz and below. On the Rx side the same process happens in reverse, but the specific frequencies used for IF1, IF2 and IF3 are slightly different. At first there is a stage that only kicks in for frequencies above 2200 MHz (bypassed otherwise), and then there is an SM4T mixer (identical to the one on Tx side) that takes in RF and Rx LO1 to produce Rx IF1. High-side injection is used, i.e., Rx LO1 is programmed to generate frequency equal to the external RF of interest PLUS the desired Rx IF1 output. Rx LO1 is programmed as follows by the instrument control software: * Rx IF1 will be at 1816.115 MHz (call it high) if the listening frequency is <= 1200 MHz or > 2200 MHz; * Rx IF1 will be at 843.085 MHz (call it low) if the listening frequency is in the intermediate range, i.e., 1200 MHz < RF <= 2200 MHz. In addition to programming Rx LO1 to produce the desired IF1 per the logic above, the software also controls switches that select one or the other IF1 filter: either the filter that passes low IF1 or the one that passes high IF1. The filters used for low and high IF1 modes are the same on both Rx and Tx sides. (The actual frequencies are slightly different, but in each case they fit within the passband of the common filter parts.) The filter for low IF1 is Murata DFC3R836P025HHD, package marking 836 CD, and the one for high IF1 is DFC31R84P075HHA, package marking CR. The two filter packages are NOT the same mechanically: the low IF1 filter is physically larger. Both parts are ceramic monoblock filters from the same family, and it seems that these filter parts were originally made for mobile phones, not for RF metrology instruments: the "836 CD" filter is for AMPS uplink band, and the "CR" filter is for DCS downlink band. On the Tx side of the board there are only two IF1 filters: one for low Tx IF1 and one for high Tx IF1. However, on the Rx side there are 3 of these ceramic filters in total: two for high IF1 (two cascaded identical filters with an amplifier in between) and just one for low IF1. Why am I covering these filters in so much detail? You probably guessed it: they are components that fail, as will be covered shortly. After the selection of either low or high IF1 filter, Rx IF1 coming out of the selected filter (either 843.085 MHz or 1816.115 MHz) is mixed with LO2, which is shared between Rx and Tx sides and fixed at 1329.6 MHz. The output of this mixer is Rx IF2 at 486.515 MHz. This Rx IF2 then passes through a pair of cascaded Sawtek 855272 filters, two identical filters with an amplifier in between, exactly the same as on the Tx side. Then there is Rx LO3 and the final mixer, producing Rx IF3 at 10.7 MHz that goes to the digital board, to the rear panel BNC output and to the WCDMA board (B68) if the latter is present. How these RXTX boards fail ========================== There are 3 specific components on this RXTX board that have been seen to fail over and over in the field: * The second of the two cascaded IF2 SAW filters (Sawtek 855272) on the Tx side often fails, breaking the Tx chain (output totally gone or extremely weak) for all frequencies. Note that there are a total of 4 identical Sawtek 855272 filters on this board (2 on Rx side, 2 on Tx side), and only one of the four fails: Tx side, second filter in the cascade. * The "836 CD" filter on the Rx side is prone to failure. When it fails, the visible symptom is severe attenuation in measured Rx signal levels for input frequencies in the 1200 MHz < RF <= 2200 MHz range. Only the Rx side filter fails, not the identical one on the Tx side! * One of the two cascaded "CR" filters on the Rx side likewise fails - this time it is the first one in the cascade. The other two identical "CR" filters on the same board (the second in cascade for Rx and the one for Tx) are likewise NOT seen to fail. The root cause of all 3 component failures has been traced to galvanic corrosion caused by direct contact between these components and Eccosorb RF absorber foam. The complete RXTX board assembly consists of the traditional PCBA plus heavy metal shields on both sides; the front and back metal shield pieces are custom- made for this board, with individually shielded cavities matching different sections of the board. Some (not all) of these cavities are filled with a special black foam called Eccosorb - it is an RF absorber, presumably added to prevent these cavities from acting as parasitic oscillators. Trouble occurs when this Eccosorb foam comes into direct contact with metal surfaces of components on the board: the result is galvanic corrosion, a process that takes many years before it results in component failure. The reason why only 3 particular filter components fail is because they got the bad luck of residing in cavities with Eccosorb - the other identical components that don't fail reside in cavities without Eccosorb. We don't know how R&S allowed this design flaw to escape and remain in their sold and field-deployed products: there is the "innocent" explanation that they simply didn't notice, and there is the conspiratorial view that this slow failure mechanism is intentional as in planned obsolescense - pick your choice of hypothesis. How to repair failed boards =========================== All 3 of the failing filter components (one SAW filter part and two ceramic monoblock filter parts) are now unobtainium. However, because so many of these RXTX boards fail in exactly the same ways, our community at large is now accumulating a very substantial "graveyard" of failed boards, and here is the good news: we can make one good board out of every two failed ones. Suppose that every RXTX board in our community's collective inventory has fully failed, leaving no failure-free boards - what now? Here is the recipe for making one good RXTX board out of two fully failed ones: 1) Out of the two failed boards, choose one to be the part donor and the other to be the part recipient. 2) Take the part donor board and harvest 3 parts from it: one of the 3 Sawtek 855272 filters that aren't subject to corrosion, and the two IF1 filters (one 836 CD and one CR) from the Tx side. Tx side IF1 filters aren't in contact with Eccosorb and thus don't corrode, and 3 out of the 4 SAW filters are likewise safe - hence we expect that every "dead" RXTX board can still serve as a donor of good parts in this manner. 3) Take the part recipient board and transplant the donor parts onto it, replacing all 3 corroded filters. 4) Before putting the repaired board back into its metal casing, cover all corrosion-prone components with Kapton tape, preventing direct galvanic contact with Eccosorb - this way the newly transplanted uncorroded components won't suffer the same fate. RXTX disassembly instructions ============================= Before you can start working on an individual RXTX board, you first need to pull it out of your CMU. Disassembly instructions are provided in the official part swapper guide from R&S (which they call "service manual"), but here is the gist: * Using a Torx T20 screwdriver, remove the 4 rear feet and lift the sleeve part of the instrument case. * Remove two small Phillips screws that secure the cover over the main board cage, and lift that cover off. * Unhook all MMCX little coax connections from the RXTX board: 3 on the top side (IF3 interface) and one on the bottom (netclock input). * Loosen and remove the two semi-rigid coax pieces that connect RF between the RXTX board and the front end. In this Mother's opinion, this step is the least pleasant of all, but it is unavoidable. * After ensuring that nothing remains connected to the RXTX board on the bottom side, pull the board out from the top. Once you got the complete RXTX board assembly out, how do you extract the actual board out of the metal casing? The not-immediately-obvious answer is that you don't need to remove all of the screws, instead there are shortcuts that will save you a lot of pain: * There are two smooth thin metal plates, one on the front side of the board (facing toward the front of the CMU when installed) and one on the back side. Each is secured with a small Phillips screw. You only need to remove the one on the front side. You don't need to remove the thin metal plate from the back side of RXTX assembly - doing so will only add more clutter and loose parts to your lab bench while the board is disassembled. * Once you remove the thin metal plate from the *front* side of your RXTX assembly, you will see all of the many screws that hold together the sandwich of two heavy metal pieces with the board in the middle. These screws are Torx T8. * Put the board down on your bench so that the side that faces the front of the CMU when installed (the side with the T8 screw heads) will become the top, with the rear side becoming bottom. * Each of the T8 screws passes through thread in the top metal piece, a hole in the PCB, and then thread in the bottom metal piece. As you loosen these screws, you don't need to remove them all the way - instead loosen each screw so that its far end comes out of the thread in the bottom metal piece, but let it remain captive in the top metal piece. Letting the screws remain captive in the top metal piece will reduce bench clutter while the board is disassembled, and there is a lot less screwing and unscrewing work to be done, as there is no need to work through the thread in the top metal piece. Once you loosen all of the T8 screws, the top metal piece should lift off, leaving just the bottom metal piece and the PCBA. The bottom metal piece has two thin metal pins sticking out of it; both the PCBA and the top metal piece align on these two pins. When you lift the top metal piece (the one with the screws), the side of the board that will be immediately exposed to you is the side that faces the front of the CMU when the board is installed. It is the Rx side, and you can confirm that you are looking at the Rx side by noting that there are two "CR" filters for high IF1, as opposed to just one. And chances are, right here at this step in the disassembly process you will see the galvanic corrosion or the lead-up to it. As you lift the top metal piece from the board, look at its inside and note the many individual cavities. Also note how some of these cavities are filled with some black foam - that's the Eccosorb. And note how only some of the cavities have Eccosorb in them, not all. Now look at the ceramic IF1 filters on the Rx side of the board. The one "CR" filter that is NOT in contact with Eccosorb will be bright copper-colored (it actually is copper), whereas the two filters that are in contact with Eccosorb (one 836 CD, one CR) will often be green instead of copper-colored on their top surface - that's patinated copper! Furthermore, there will typically be some black Eccosorb material directly adhered to the corroding top surfaces of those two unlucky filters. Now lift the PCBA off the two metal pins, separating it from the bottom metal piece. Like you did with the top metal piece, observe the inside of the bottom metal piece: note which cavities have Eccosorb in them and which don't. Then flip the board over and look at its Tx side. You will see that there are only two ceramic IF1 filters on this side (one 836 CD and one CR), and both should be in pristine shape, bright copper-colored, no corrosion - these two are not in contact with Eccosorb! Now look at the two Sawtek 855272 filters on the Tx side. The one closer to the middle of the board will often appear in worse physical condition that the other 3 - and the culprit is once again in contact with Eccosorb. MACOM SM4T mixer corrosion ========================== Neither I nor my collaborator on this project have seen an RXTX board on which either the Rx SM4T mixer or the Tx one went bad - i.e., we haven't seen a failure in this part *yet*. However, this mixer *is* in contact with Eccosorb, and looking visually at the collection of RXTX boards in my possession, I (Mother Mychaela) see definite signs of corrosion - the metal surface of this SM4T mixer component is beginning to corrode. Therefore, as a preventative measure, I recommend cleaning off any Eccosorb that is adhered to this component and then covering the component with Kapton tape before putting the board back into its metal casing. Unlike the failing filters, this MACOM SM4T mixer is still available new - but it's an expensive component, so let's protect these mixers from corrosion.