# This file is an EXAMPLE of the cable configuration file for fc-cmu200d.
# The cable config file describes the insertion loss of the cabling
# between the CMU200 and the DUT for each uplink and downlink frequency
# band; knowing exactly what this cable loss amounts to is very critical
# for proper calibration; simply ignoring or neglecting this cable loss
# or putting in "off the wall" numbers is *not* an acceptable option.
#
# The numbers given in this EXAMPLE file have a chance of being correct
# *only* for the specific cabling setup which the Mother of FreeCalypso
# currently uses in her lab; if you are setting up your own calibration
# station, you need to replace them with your own numbers that are
# correct for YOUR cabling setup.

cable-loss 850ul 1.5
cable-loss 850dl 1.5
cable-loss 900ul 1.5
cable-loss 900dl 1.6

cable-loss 1800ul 2.2
cable-loss 1800dl 2.2
cable-loss 1900ul 2.3
cable-loss 1900dl 2.3